PURPOSE: Providing correct rotational alignment of femoral component in total knee arthroplasty (TKA) is mandatory to achieve correct kinematics, good ligament balance and patellar tracking. The purpose of this study was to evaluate potential clinical applications of functional flexion axis (FFA) by analysing the differences between pre- and post-implant placement. This evaluation was supported by the analysis of repeatability, assessing the robustness of the proposed method. METHODS: Anatomical acquisitions and passive kinematics were acquired on 87 patients undergoing TKA using a commercial navigation system. Knee FFA was estimated, before and after implant positioning, from three flexion-extension movements between 0 and 120°. The angle between FFA and transepicondylar axis was analysed in frontal and axial planes. Repeatability coefficient and intraclass correlation coefficient (ICC) were used to analyse the reliability and the agreement in identifying the axis. RESULTS: The analysed angle presented differences between pre- and post-operative conditions only in the frontal plane (from -8.3 ± 5.5° to -2.8 ± 5.3°) (p < 0.0001). There was good intraobserver reliability and agreement. Repeatability coefficient ranged between 4.4° (3.7-4.9°) and 3.4° (2.9-3.8°), the ICC between 0.87 (0.83-0.91) and 0.93 (0.90-0.95) and the standard deviation ranged between 1.3 and 1.0°. CONCLUSIONS: The present study demonstrated that TKA affected the estimation of FFA only in the frontal plane. This method reported good repeatability, demonstrating its usefulness for clinical purposes particularly to evaluate rotational positioning of the femoral component in the axial plane. LEVEL OF EVIDENCE: Case series, Level IV.

Analysis of knee functional flexion axis in navigated TKA: Identification and repeatability before and after implant positioning

MARCACCI, MAURILIO
2014-01-01

Abstract

PURPOSE: Providing correct rotational alignment of femoral component in total knee arthroplasty (TKA) is mandatory to achieve correct kinematics, good ligament balance and patellar tracking. The purpose of this study was to evaluate potential clinical applications of functional flexion axis (FFA) by analysing the differences between pre- and post-implant placement. This evaluation was supported by the analysis of repeatability, assessing the robustness of the proposed method. METHODS: Anatomical acquisitions and passive kinematics were acquired on 87 patients undergoing TKA using a commercial navigation system. Knee FFA was estimated, before and after implant positioning, from three flexion-extension movements between 0 and 120°. The angle between FFA and transepicondylar axis was analysed in frontal and axial planes. Repeatability coefficient and intraclass correlation coefficient (ICC) were used to analyse the reliability and the agreement in identifying the axis. RESULTS: The analysed angle presented differences between pre- and post-operative conditions only in the frontal plane (from -8.3 ± 5.5° to -2.8 ± 5.3°) (p < 0.0001). There was good intraobserver reliability and agreement. Repeatability coefficient ranged between 4.4° (3.7-4.9°) and 3.4° (2.9-3.8°), the ICC between 0.87 (0.83-0.91) and 0.93 (0.90-0.95) and the standard deviation ranged between 1.3 and 1.0°. CONCLUSIONS: The present study demonstrated that TKA affected the estimation of FFA only in the frontal plane. This method reported good repeatability, demonstrating its usefulness for clinical purposes particularly to evaluate rotational positioning of the femoral component in the axial plane. LEVEL OF EVIDENCE: Case series, Level IV.
2014
CAS
Knee functional flexion axis
Repeatability
TKA
Aged
Aged
80 and over
Arthroplasty
Replacement
Knee
Biomechanical Phenomena
Femur
Humans
Knee Joint
Knee Prosthesis
Middle Aged
Osteoarthritis
Knee
Range of Motion
Articular
Reproducibility of Results
Torsion
Mechanical
Treatment Outcome
Surgery
Computer-Assisted
Orthopedics and Sports Medicine
Surgery
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11699/31792
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 10
social impact