: For the diagnosis of anti-MAG polyneuropathy the commercial ELISA manufacturer currently recommends a cut-off of 1000 Bühlmann Titer Units (BTU). We analyzed sera from 80 anti-MAG neuropathy patients and 383 controls (with other neuropathies or healthy controls) to assess the ELISA sensitivity and specificity at different thresholds. A better combination of sensitivity/specificity was found at a threshold >1500 BTU than at >1000 BTU. The best value of specificity was obtained at threshold >7000 BTU. There was a diagnostic grey area between 1500 and 7000 BTU in which the clinical phenotypes as well as electrophysiological studies need to be carefully assessed particularly to differentiate CIDP and anti-MAG neuropathy.
Sensitivity and specificity of a commercial ELISA test for anti-MAG antibodies in patients with neuropathy
Doneddu, Pietro Emiliano;
2020-01-01
Abstract
: For the diagnosis of anti-MAG polyneuropathy the commercial ELISA manufacturer currently recommends a cut-off of 1000 Bühlmann Titer Units (BTU). We analyzed sera from 80 anti-MAG neuropathy patients and 383 controls (with other neuropathies or healthy controls) to assess the ELISA sensitivity and specificity at different thresholds. A better combination of sensitivity/specificity was found at a threshold >1500 BTU than at >1000 BTU. The best value of specificity was obtained at threshold >7000 BTU. There was a diagnostic grey area between 1500 and 7000 BTU in which the clinical phenotypes as well as electrophysiological studies need to be carefully assessed particularly to differentiate CIDP and anti-MAG neuropathy.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.