Drug-eluting stents (DES) have made a tremendous impact on the practice of percutaneous coronary intervention. Recently however, long-term DES failures have become a focal point, particularly with restenosis and thrombosis. An uncommon, yet important cause of DES failure is stent fracture. Of the two established first generation DES, the sirolimus-eluting stent (SES) has been particularly linked to cases of stent fracture, likely as a result of its closed cell design compared with other DES employing an open cell system. We present 2 cases of SES fracture confirmed using high-resolution intravascular optical coherence tomography giving unique insights into the in-vivo appearance of this complication. (C) 2008 Published by Elsevier Ireland Ltd.
The use of intra-coronary optical coherence tomography for the assessment of sirolimus-eluting stent fracture
Ferrante G;
2009-01-01
Abstract
Drug-eluting stents (DES) have made a tremendous impact on the practice of percutaneous coronary intervention. Recently however, long-term DES failures have become a focal point, particularly with restenosis and thrombosis. An uncommon, yet important cause of DES failure is stent fracture. Of the two established first generation DES, the sirolimus-eluting stent (SES) has been particularly linked to cases of stent fracture, likely as a result of its closed cell design compared with other DES employing an open cell system. We present 2 cases of SES fracture confirmed using high-resolution intravascular optical coherence tomography giving unique insights into the in-vivo appearance of this complication. (C) 2008 Published by Elsevier Ireland Ltd.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.